• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOMATOM DRIVE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOMATOM DRIVE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 10431700
Medical Device Problem Code Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The initial reporter name was not provided to siemens as of the date of this report and have been requested by siemens.A siemens employee reported the complaint to siemens.Siemens has initiated a technical investigation of the reported event.A root cause has not been identified.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that during a dual energy ct contrast scan of a (b)(6) male patient using the somatom drive system, the runoff scan could not be loaded after monitoring and the examination was aborted.The scan had to be repeated using a different ct system and additional contrast media was administered.The second low dose overview scan (topogram) was completed successfully.The patient did not suffer any adverse health issues due to the delay in the examination, additional low dose x-ray exposure and additional contrast media.This report has been submitted with an abundance of caution.
 
Additional Manufacturer Narrative
Siemens performed technical evaluation of the reported event and system.The issue was caused by the customized scan protocol "de_runoff", in dual energy spiral range "de runoff" with rottime 1.0s and coll.Slice 64x0.6.The range is based on siemens mode de_extrangio_bonerem_ir.Adult", which is designed with rottime 0.33s, coll.Slice 32x0.6.The scanner could not find the correct calibration table with the invalid parameter which resulted in the reported problem.This system behavior will be resolved in the next servicepack vb20a_sp5.A defect for follow-up in siemens change management has been opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM DRIVE
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
MDR Report Key11640518
Report Number3004977335-2021-72253
Device Sequence Number3049289
Product Code JAK
UDI-Device Identifier04056869006970
UDI-Public04056869006970
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K190578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10431700
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/12/2021
Supplement Date Received by Manufacturer04/19/2021
Initial Report FDA Received Date04/09/2021
Supplement Report FDA Received Date04/23/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age10 YR
Patient Weight82
-
-