| Model Number |
A3219 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident device is anticipated to return.A follow-up report will be provided upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: cags.[name] was scrubbed for the case one of the clamps that has the inserts that they put on the jaws.[name] applied the inserts to the jaws and did one click of the ratchet as she normally does and put it on her mayo.When she went to give it to the surgical team they noticed that the ratchet wasn¿t working.They looked at the box joint of the clamp and noticed a fracture through one of the arms.This same issue has happened in january with another 1 of 6 clamps that were purchased at the same time.Intervention: ni.Patient status: patient is fine, and completely unaffected.
|
| |
|
Event or Problem Description
|
|
Procedure performed: cags.[name] was scrubbed for the case one of the clamps that has the inserts that they put on the jaws.[name] applied the inserts to the jaws and did one click of the ratchet as she normally does and put it on her mayo.When she went to give it to the surgical team they noticed that the ratchet wasn¿t working.They looked at the box joint of the clamp and noticed a fracture through one of the arms.This same issue has happened in january with another 1 of 6 clamps that were purchased at the same time.Intervention: ni.Patient status: patient is fine, and completely unaffected.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed that the box lock of the returned unit was fractured.Corrosion was also observed around the location of the break.Based on the condition of the returned unit, the fracture was caused by the corrosion that was observed around the location of the break.Reusable devices are designed to withstand repeated use and sterilization cycles.However, the lifespan of the device is dependent on the storage, cleaning, and usage conditions.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
| |
|