SMITH & NEPHEW, INC. 3.2MM T-HANDLE TROCAR; PIN, FIXATION, THREADED Back to Search Results
Model Number 71674074
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery, the handle broke of from the body of the t-handle trocar, while was being used outside the patient.The procedure was finished using a backup device from smith and nephew, without surgical delay and no injury to the patient.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device broke into two pieces at the handle component.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.2MM T-HANDLE TROCAR
Common Device Name
PIN, FIXATION, THREADED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11642772
Report Number1020279-2021-02807
Device Sequence Number14242309
Product Code JDW
UDI-Device Identifier03596010560704
UDI-Public03596010560704
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K120871
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71674074
Device Catalogue Number71674074
Device Lot Number11HCT0007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2021
Initial Date Received by Manufacturer 03/15/2021
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date04/09/2021
Supplement Report FDA Received Date05/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-