• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS ELITE; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS ELITE; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 10673
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2021
Event Type  Injury  
Manufacturer Narrative
Age at time of event: above 18 years and older.
 
Event Description
It was reported that stent fracture occurred.Vascular access was obtained via femoral artery.The 80% stenosed target lesion containing a less than equal to 45 degrees bend was located in the mildly tortuous and mildly calcified left circumflex artery.After the lesion was pre-dilated with a semi-compliant balloon catheter, a 20 x 3.00 promus elite drug-eluting stent was implanted.However, during post dilatation, angiography appeared to show the stent had fractured.Another device was advanced to complete the procedure.No patient complications were reported and the patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS ELITE
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11644019
MDR Text Key244845130
Report Number2134265-2021-04004
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/18/2022
Device Model Number10673
Device Catalogue Number10673
Device Lot Number0025944752
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-