• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSTEC BARBADOS INC. SOFTEC HD; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSTEC BARBADOS INC. SOFTEC HD; INTRAOCULAR LENS Back to Search Results
Model Number SOFTECHD+23.0
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A full audit of all batch documentation relating to the production of the device was performed.The audit concluded that all procedures in manufacturing and packaging of the device had been conducted correctly.Batch reconciliation was 100.0%.Based on the assessment of our batch documentation, lenstec can confirm that all procedures in the manufacturing and packaging of the lens were conducted correctly.
 
Event or Problem Description
Lenstec received an email stating " lens implanted (b)(6) 2020 and explanted on (b)(6) 2020.Lenstec received a form indicating 'contaminants inside'.A back up lens was implanted and there was no patient injury.The specific model of th einjector, cartridge and viscoelastic solution was not provided.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC HD
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB 
MDR Report Key11645213
Report Number9613160-2021-00010
Device Sequence Number11553370
Product Code HQL
UDI-Device Identifier00844369027410
UDI-Public00844369027410
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/09/2025
Device Model NumberSOFTECHD+23.0
Device Catalogue NumberSOFTEC HD
Device Lot Number200285
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2021
Initial Report FDA Received Date04/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-