• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH SLIDING CORE, UHMPWE,8MM; PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH SLIDING CORE, UHMPWE,8MM; PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED Back to Search Results
Model Number 400-142F
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.Device was lost or discarded by the customer.
 
Event or Problem Description
The following event was reported:" patient presented at dr.(b)(6).Office with ankle instability.He examined the joint and determined that a new poly was necessary.We performed the revision surgery and swapped the 8mm poly for a 9mm poly.Original dos (b)(6) 2020.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
The following event was reported:" patient presented at dr.M.Office with ankle instability.He examined the joint and determined that a new poly was necessary.We performed the revision surgery and swapped the 8mm poly for a 9mm poly.Original dos (b)(6) 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIDING CORE, UHMPWE,8MM
Common Device Name
PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key11645428
Report Number0008031020-2021-00145
Device Sequence Number3049889
Product Code NTG
UDI-Device Identifier00886385026947
UDI-Public00886385026947
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
P050050
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number400-142F
Device Catalogue Number400142F
Device Lot Number1543203
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/31/2021
Supplement Date Received by Manufacturer06/18/2021
Initial Report FDA Received Date04/09/2021
Supplement Report FDA Received Date06/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
-
-