Model Number 436-28-000
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 03/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer narrative: additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
2nd revision surgery due to revised an onkos proximinal femur.The locking ring was locked, but there was a dislocation, a new product styled proximal body was installed.Doctor felt the newer body would solve the issue.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as dislocation.The previous surgery and the surgery detailed in this event occurred 10 days apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this compliant was a revision surgery due to dislocation.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions, improper surgical technique, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Second revision surgery - revised an onkos proximal femur.The locking ring was locked, but there was a dislocation, a new product styled proximal body was installed.Surgeon felt the newer body would solve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTRAINED LINER HIP
Common Device Name
LINER, LOCKING RING, CONSTRAINED
Manufacturer (Section D)
ENCORE MEDICAL L. P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key11645541
Report Number1644408-2021-00232
Device Sequence Number3496962
Product Code KWZ
UDI-Device Identifier00888912085663
UDI-Public(01)00888912085663
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K023794
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/24/2026
Device Model Number436-28-000
Device Catalogue Number436-28-000
Device Lot Number496C1089
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/11/2021
Supplement Date Received by Manufacturer04/22/2021
Initial Report FDA Received Date04/09/2021
Supplement Report FDA Received Date05/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
436-28-010, LOT 502C1008
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age66 YR
-
-