| Model Number |
FRED4007-PMA |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Intracranial Hemorrhage (1891); Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440)
|
| Date of Event |
03/10/2021
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and not returned to the manufacturer for analysis.Additionally, procedural images were not provided; therefore, the reported event cannot be confirmed.The instructions for use identifies death and device migration as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a fred 27 stent was placed for the treatment of an unruptured anterior choroidal aneurysm.The fred stent was deployed successfully and appeared to have good wall apposition.On a final angiographic run, the fred was observed to be positioned roughly 2-3mm distal into the middle cerebral artery (mca) with slow flow.Additional angiographic runs revealed an occlusion of the mca due to thrombosis of the vessel.The physician attempted to remove and recapture the fred.An intraoperative bleed occurred during the manipulations within the blood vessel and the physician then coiled the ica to stop it.An external ventricular drain was placed to drain blood.The patient later passed away.
|
| |
|
Additional Manufacturer Narrative
|
|
A4: 60kg.B5: additional information indicated that the patient had abnormal anatomy secondary to the patient's previous subarachnoid hematoma.
|
| |
|
Search Alerts/Recalls
|