| Model Number |
66800900 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during inspection upon removing the backing on this dressing a significant amount of the adhesive on the pad comes off onto the backing.No patient was involved.
|
| |
|
Additional Manufacturer Narrative
|
|
We have now completed our investigation into the reported complaint.The device intended to be used in treatment has been returned and evaluated establishing a relationship with the reported event.The visual inspection found when the plastic handle was peeled off dressing, silicone offset onto the plastic handle.The functional evaluation confirmed that the dressing was less adhesive.A wound contact layer raw material quality issue has been identified as the root cause.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.The complaint history file contains further instances of the reported events.An ongoing internal action is being carried out into the reported failure mode to reduce further instances of the reported event, therefore no further actions are deemed necessary into this specific complaint.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
|
| |
|