• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MEDIAL UTILITY SCD INSERT; TRAY, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MEDIAL UTILITY SCD INSERT; TRAY, SURGICAL, INSTRUMENT Back to Search Results
Model Number 74036176
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, while inspecting instruments, it was noticed that this tray is bent and lid does not set well on it.No case involved.
 
Additional Manufacturer Narrative
H6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The lid on the device was found to be bent , rendering the device inoperable.The device shows signs of extensive use.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDIAL UTILITY SCD INSERT
Common Device Name
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11646831
Report Number1020279-2021-02834
Device Sequence Number3072915
Product Code FSM
UDI-Device Identifier00885556707005
UDI-Public00885556707005
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74036176
Device Catalogue Number74036176
Device Lot Number20EAV0019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2021
Initial Date Received by Manufacturer 03/18/2021
Supplement Date Received by Manufacturer09/08/2021
Initial Report FDA Received Date04/11/2021
Supplement Report FDA Received Date09/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-