((b)(4)) it was reported that during inspection process, it was found that the threads on the lag screw driver are bent.There was no case involved.((b)(4)) it was reported that during inspection process, it was found that the trigen t-handle does not capture and release driver properly.There was no case involved.
|
|
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device did not confirm the stated failure mode.The device was found to have scratches and nicks, rendering the device inoperable.The device was manufactured in 2018 and shows signs of extensive use.A functional evaluation of the returned device confirmed the stated failure mode.The device was paired with the mating part but could not function, rendering the device inoperable.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|