• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS SMALL T8 TAPERED DR SHAFT - SHORT; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS SMALL T8 TAPERED DR SHAFT - SHORT; NAIL, FIXATION, BONE Back to Search Results
Model Number 71175078
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use, for a evos small t8 tapered dr shaft - short it was noted that the driver has become rounded and does not capture screw.No other complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection was performed and confirmed theevos small t8 tapered dr shaft ¿ short is rounded off and damaged.The device shows significant signs of wear/usage.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL T8 TAPERED DR SHAFT - SHORT
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11646838
Report Number1020279-2021-02841
Device Sequence Number3084441
Product Code JDS
UDI-Device Identifier00885556636084
UDI-Public00885556636084
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175078
Device Catalogue Number71175078
Device Lot Number19DUP0004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2021
Initial Date Received by Manufacturer 03/17/2021
Supplement Date Received by Manufacturer06/07/2021
Initial Report FDA Received Date04/11/2021
Supplement Report FDA Received Date06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-