SMITH & NEPHEW, INC. JOURNEY PAT CEM CLAMP; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY Back to Search Results
Model Number 74019801
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2021
Event Type  malfunction  
Event Description
It was reported that, for a journey pat cem clamp it was noted during processing that the red o-ring was missing from the device.No case reported; therefore, there was no patient involvement.
 
Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY PAT CEM CLAMP
Type of Device
PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11646910
MDR Text Key244855381
Report Number1020279-2021-02869
Device Sequence Number1
Product Code MBH
UDI-Device Identifier03596010557339
UDI-Public03596010557339
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74019801
Device Catalogue Number74019801
Device Lot Number14EM00641
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/16/2021
Initial Date FDA Received04/11/2021
Supplement Dates Manufacturer Received05/16/2021
Supplement Dates FDA Received05/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-