Model Number 71420526
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during tka surgery the surgeon attempted to implant the gns ii con ins sz3-4 13mm and was unable to get it to seat in the tibial base.The procedure was completed without delay using a smith and nephew back-up device.No other complications were reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II CON INS SZ3-4 13MM
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11646943
Report Number1020279-2021-02900
Device Sequence Number3367783
Product Code JWH
UDI-Device Identifier03596010208057
UDI-Public03596010208057
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71420526
Device Catalogue Number71420526
Device Lot Number19DM07007
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/16/2021
Supplement Date Received by Manufacturer07/21/2021
Initial Report FDA Received Date04/11/2021
Supplement Report FDA Received Date07/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-