• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIO; UNO INSET II 60/6 PCC PINK MIMX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIO; UNO INSET II 60/6 PCC PINK MIMX Back to Search Results
Model Number MMT-923
Medical Device Problem Codes Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in the united states.On (b)(6) 2021, it was reported that while unwinding the infusion set's tubing before insertion, it accidently pulled out the set at the tubing connector and this issue occurred a couple of weeks ago.Currently, patient's blood glucose level was 152 mg/dl.Reportedly, there was no stress or pull on the tubing and the pump was not dropped with the set connected to patient's body.No further information available.
 
Event or Problem Description
On 12-may-2021: follow up information was submitted to update the lot number, expiry date and awareness date.Moreover, the result of complaint investigation of the returned used device (2 sets) showed that all the test results were within specification.Based on the result: medical device problem code, component code, type of investigation code, investigation findings code and investigation conclusions code were updated.Unomedical reference number (b)(4).Event occurred in the united states.On (b)(6) 2021, it was reported that while unwinding the infusion set's tubing before insertion, it accidently pulled out the set at the tubing connector and this issue occurred a couple of weeks ago.Currently, patient's blood glucose level was 152 mg/dl.Reportedly, there was no stress or pull on the tubing and the pump was not dropped with the set connected to patient's body.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIO
Common Device Name
UNO INSET II 60/6 PCC PINK MIMX
MDR Report Key11647334
Report Number3003442380-2021-00207
Device Sequence Number3657863
Product Code FPA
UDI-Device Identifier05705244007697
UDI-Public05705244007697
Combination Product (Y/N)Y
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date06/01/2023
Device Model NumberMMT-923
Device Lot Number5323777
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2021
Supplement Date Received by Manufacturer05/05/2021
Initial Report FDA Received Date04/12/2021
Supplement Report FDA Received Date05/12/2021
Patient Sequence Number1
-
-