Model Number SK15014
Medical Device Problem Codes Break (1069); Retraction Problem (1536)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Expiry date: 04/2022.
 
Event or Problem Description
It was reported that while pulling the catheter over the wire which were in a stenosed and fully calcified anterior tibial artery (ata), resistance was suddenly gone and the proximal part of the catheter was allegedly found to be torn inside the ata.The device was not recovered and the procedure was completed using another device.There was no reported patient injury.
 
Event or Problem Description
It was reported that while pulling the catheter over the wire which were in a stenosed and fully calcified anterior tibial artery (ata), resistance was suddenly gone and the proximal part of the catheter was allegedly found to be torn inside the ata.The device was not recovered and the procedure was completed using another device.Patient current status is unknown.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the alleged retraction problem and catheter break issue could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Expiry date: 04/2022.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEEKER
Common Device Name
CROSSING SUPPORT CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11647821
Report Number2020394-2021-00839
Device Sequence Number14373474
Product Code DQY
UDI-Device Identifier00801741067907
UDI-Public(01)00801741067907
Combination Product (Y/N)N
Initial Reporter CountryQA
PMA/510(K) Number
K103367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSK15014
Device Catalogue NumberSK15014
Device Lot NumberVTER0497
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/14/2021
Supplement Date Received by Manufacturer05/31/2021
Initial Report FDA Received Date04/12/2021
Supplement Report FDA Received Date06/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-