Patient admitted 3 days after pacemaker generator change with syncope.Tele in er showed intermittent inhibition of pacing with prolonged pauses.The new pacer had malfunctioned and switched to "safety core" mode.The included vvi mode with sensitivity 0.25 mv unipolar which resulted in far field and/or myopotential sensing and inhibited ventricular output.A temporary place was placed.The permanent pacemaker generator was replaced.The ventricular lead was found to have an insulated crack at the pin insertion point though electrical measurements were normal.New right ventricular pacing lead was implanted along with a new generator.Manufacturer response for accolade mri (b)(4), (brand not provided) (per site reporter) the product was given to boston scientific for evaluation.
|