| Model Number |
DEPTH GUIDE, MINI |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the top part of the depth guide fell off when the surgeon went to measure the drill hole for an implant.The tip was removed from the patient and the distal radius fracture case was completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, the tip was found to be detached from the inner component.Evaluation revealed the weld broke but the tip itself is whole.The tip was also found to be bent.A likely cause of the event is prying/leveraging the device during use.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, the tip was found to be detached from the inner component.Evaluation revealed the weld broke but the tip itself is whole.The tip was also found to be bent.A likely cause of the event is prying/leveraging the device during use.
|
| |
|