Model Number 4400804S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be sent in upon 30 days after receipt.
 
Event or Problem Description
As reported, the 8mm x 4cm x 80cm powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (bc) was used, but it ruptured within its nominal pressure.It was replaced with another balloon catheter (unknown) and the procedure was completed.There was no reported patient injury.The lesion was the common iliac artery.The device will not be returned for analysis as it was discarded.Additional procedural details were requested but are unknown.
 
Additional Manufacturer Narrative
As reported, the 8mm x 4cm x 80cm powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (bc) was used, but it ruptured within its nominal pressure.It was replaced with another balloon catheter (unknown) and the procedure was completed.There was no reported patient injury.The lesion was the common iliac artery.Additional procedural details were requested but are unknown.The product was not returned for analysis as it was discarded.A product history record (phr) review of lot 82200840 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown, may have contributed to the reported event.However, with the limited amount of information provided and without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the safety information in the instructions for use, ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 8MM4CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CARDINAL HEALTH MEXICO
av prado sur 150 2d0 piso
ciudad de mexico 11000
MX  11000
MDR Report Key11648209
Report Number9616099-2021-04396
Device Sequence Number3657755
Product Code LIT
UDI-Device Identifier20705032059814
UDI-Public20705032059814
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model Number4400804S
Device Catalogue Number4400804S
Device Lot Number82200840
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/16/2021
Supplement Date Received by Manufacturer04/14/2021
Initial Report FDA Received Date04/12/2021
Supplement Report FDA Received Date04/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-