• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TRM-560
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Swollen Lymph Nodes/Glands (4432)
Date of Event 02/22/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Device evaluation: visual analysis of the returned device identified it to be underweight, a broken shell and non-penetrating nicks.A microscopic analysis was performed which identified a sharp edge with non-penetrating nicks, assessed as a surgical impact opening on the posterior side.A dimension measurement in the shell was performed which identified the thickness within specification.Based on the device analysis the final assessment is: a sharp broken edge with non-penetrating nicks due to surgical impact.Also observed were non-penetrating nicks.The event of lymphadenopathy is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture, lymphadenopathy.Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.
 
Event or Problem Description
Patient reported left side "dents and wrinkling", exchange from textured to smooth implants due to the patient's concern with the product, "itching and bumps in her armpit", rupture, pain and "lopsided".Patient additionally reported "cyst and that it is nothing to do with the implants" which is not device related.Healthcare professional reported left side rupture and exchange from a textured to smooth breast implant due to the patient¿s concern with the product.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
MDR Report Key11648543
Report Number9617229-2021-06483
Device Sequence Number11676883
Product Code FTR
UDI-Device Identifier10888628008175
UDI-Public10888628008175
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/05/2020
Device Model NumberTRM-560
Device Catalogue NumberTRM-560
Device Lot Number2784364
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/14/2019
Initial Report FDA Received Date04/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
-
-