| Model Number |
26924 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the tip fractured.A 6x40x130 innova self-expanding stent was selected for use to treat arterial stenosis in the left leg.Upon removal from the packaging a fracture to the distal tip was observed.The catheter sheath was noted to be moving freely.A new stent, same model, was opened and successfully delivered.There were no patient complications.
|
| |
|
Event or Problem Description
|
|
It was reported the tip fractured.A 6x40x130 innova self-expanding stent was selected for use to treat arterial stenosis in the left leg.Upon removal from the packaging a fracture to the distal tip was observed.The catheter sheath was noted to be moving freely.A new stent, same model, was opened and successfully delivered.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Device eval by mfr: the returned product consisted of an innova self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the nosecone is separated.Microscopic examination revealed no additional damages.Inspection of the remainder of the device, revealed no other damage or irregularities.
|
| |
|