| Catalog Number |
7210569 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the positioner spider limb was slipping/moving when under pressure before a shoulder scope.The procedure was completed with a working spider unit.No injuries, delay or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product.No issues were noted.A functional evaluation was performed.The device failed the weight test.The piston migration was at 2.3 inches.The reported failure was confirmed and the root cause has been associated with a seal failure.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
| |
|
Search Alerts/Recalls
|