| Model Number |
TMICL13.2 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation type 4110: lens work order search-one similar complaint type event reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens, -12.5/+1.0/055 (sphere/cylinder/axis), into the patient's eye.Reportedly, the lens was absent of the markings on the haptics to verify the correct orientation.The lens was explanted and tore upon removal.See mfr rep# 2023826-2021-00956 for replacement lens.Attempts to obtain additional information have not been successful.
|
| |
|
Additional Manufacturer Narrative
|
|
B1-corrected b5-previously stated the lens was implanted.The doctor "felt iffy about implanting it" so the back-up lens was implanted instead.H6-health impact code corrected.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3-device evaluation: the lens was returned in liquid in a vial.Visual inspection found that the haptic was torn.Haptic markings are noted.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no similar complaint was reported for units within the same lot.Claim#: (b)(4).
|
| |
|