| Model Number |
72200561 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that there was not light on the camera head.Incident occurred while setting-up to the procedure.A back-up device was available and no delays occurred.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was returned to the designated complaint station for independent evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device did not identify any issues.Functional evaluation revealed that there was yellow noise in the video output when the connector is moved.The complaint has been confirmed and the root cause has been associated with an electrical component failure.Factors, unrelated to the design or manufacture of the device which could have contributed to the complaint event, include a damaged connector.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
| |
|