• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQ7594
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the conquest pta dilatation catheter products that are cleared in the us.The pro code and 510 k number for the conquest pta dilatation catheter products are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 09/2022).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the proximal fiber material of the balloon was allegedly misaligned after inflation.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the conquest pta dilatation catheter products that are cleared in the us.The pro code and 510 k number for the conquest pta dilatation catheter products are identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one conquest pta dilatation catheter was returned for the evaluation.Visual evaluation was performed and device appeared to be bloody.During the microscopic evaluation fiber disturbance was noted to the proximal end of the balloon.No other anomalies were noted to the device.Functional test was not able to performed due to the nature of the balloon.Therefore, the investigation is confirmed for the reported material frayed issue as fiber disturbance was noted during the visual evaluation.A definitive root cause for the material frayed issue could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 09/2022), g3, h6 (method).H11: h6 (result and conclusion).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the proximal fiber material of the balloon was allegedly misaligned after inflation.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11654398
Report Number2020394-2021-00843
Device Sequence Number3520425
Product Code DQY
UDI-Device Identifier00801741063190
UDI-Public(01)00801741063190
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ7594
Device Catalogue NumberCQ7594J
Device Lot NumberREDX5096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2021
Initial Date Received by Manufacturer 03/17/2021
Supplement Date Received by Manufacturer07/02/2021
Initial Report FDA Received Date04/13/2021
Supplement Report FDA Received Date07/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-