MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARTICULAR SURFACE WITH SEGMENTAL HINGE POST SIZE D 14 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARTICULAR SURFACE WITH SEGMENTAL HINGE POST SIZE D 14 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2021-01042.
 
Event or Problem Description
It was reported that a patient underwent a left knee arthroplasty on an unknown date.Subsequently the patient was revised due to a broken hinge mechanism and post on an rhk.Attempts have been made and no additional information has been provided.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
It was reported that the patient underwent initial total knee arthroplasty with unknown devices.Two years post initial surgery, the patient was implanted with zimmer biomet devices with hinge.Hinge failed a year later.There was no specific trauma, patient was just getting up from the chair.Hinge was revised to one that the surgeon felt was more durable.Surgeon states between the patient's tendency to hyperextend and her muscle weakness, it puts an extra load on the hinge mechanism.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE WITH SEGMENTAL HINGE POST SIZE D 14 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11654752
Report Number0001822565-2021-01041
Device Sequence Number14354012
Product Code KRO
UDI-Device Identifier00889024195875
UDI-Public(01)00889024195875(17)260430(10)64007194
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K070978
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Model NumberN/A
Device Catalogue Number00585004014
Device Lot Number64007194
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/18/2021
Supplement Date Received by Manufacturer06/08/2021
06/15/2021
Initial Report FDA Received Date04/13/2021
Supplement Report FDA Received Date06/14/2021
06/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ITEM 00585007514 LOT 63980603.; ITEM 00585007514 LOT 63980603
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight66
-
-