U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-FEM-CELECT
Device Problems Inadequate Filtration Process (2308); Obstruction of Flow (2423); Structural Problem (2506); Device Tipped Over (2589); Appropriate Term/Code Not Available (3191)
Patient Problems Pulmonary Embolism (1498); Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Thrombosis/Thrombus (4440); Insufficient Information (4580)
Event Type  Injury  
Manufacturer Narrative
It has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog # is unknown but referred to as celect pt.Occupation: non-healthcare professional.Catalog number and lot number are unknown, however, the alleged celect pt is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
Original: it is alleged that the patient received a celect platinum (pt) inferior vena cava (ivc) filter on (b)(6) 2013, and the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Event Description
Patient allegedly received an implant on (b)(6) 2013 via the right common femoral vein due to deep vein thrombosis and pulmonary embolism.Patient is alleging tilt, vena cava perforation.Patient further alleges pain in lower left abdominal region, physical limitations.Report from computerized tomography (ct): "the apex of the ivc filter appears to lie against the anterior ivc wall.There may be mild ivc filter tilting.The apex of the ivc filter appears to lie 11 mm below the left renal vein.The filter legs perforate anteriorly and posteriorly from the ivc.Anteriorly the legs perforate approximately 14 mm while posteriorly the legs penetrate approximately 30 mm extending adjacent to the anterior l3-4 disc space.".
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: investigation investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.New pe as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: pulmonary embolism.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain and physical limitations are directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.No relevant notes on wo for device.One other complaint ((b)(4), different issue, celect, vena cava perforation, embedment (difficult to retrieve), anxiety, mental anguish, stress) on lots.Product is manufactured and inspected according to current controls.The following allegations have been investigated: pulmonary embolism (pe), vena cava (vc)/organ perforation, deep vein thrombosis (dvt), tilt, pain, physical limitations.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.The following fields were updated per additional information received: a2, a4, b1, b5, b6, b7, d1, d4, d6a, h4, h6.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT NAVALIGN FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
MDR Report Key11655450
MDR Text Key245867809
Report Number3002808486-2021-01080
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/07/2015
Device Catalogue NumberIGTCFS-65-1-FEM-CELECT
Device Lot NumberE3005025
Was Device Available for Evaluation? No
Date Manufacturer Received05/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight83
-
-