ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)
|
Back to Search Results |
|
Catalog Number 8065752180 |
Device Problem
Suction Failure (4039)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that there was no aspiration during the phaco step of a cataract procedure; tubing blockage is suspected.The product was replaced and the procedure was completed.There was no patient harm.
|
|
Manufacturer Narrative
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.In-process controls are established to ensure each final assembled cassette is verified that all required tests have been performed and all acceptance criteria are met prior to release.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
The product was returned and the sample was visually inspected and functionally tested.No defects were observed.Both irrigation and aspiration luers were intact.The sample primed and tuned with a system handpiece successfully.The sample could be recognized and the service data could be retrieved from the console.Maximum vacuum, aspiration, and irrigation were tested at appropriate settings and met specifications.The root cause of the customer's complaint could not be established since the returned sample met specifications.No contributing factors could be identified that could cause the reported complaint.After the investigation of this complaint, it has been determined that this sample met specifications.Therefore, no action will be taken at this time.Quality assurance will continue to monitor customer complaints via the complaint review meetings, and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).
|
|