• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COV LP/LLC, AUGUSTA ILA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COV LP/LLC, AUGUSTA ILA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 3971
Medical Device Problem Codes Break (1069); Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during open abdominal surgery, the tissue was stapled, but the cutter did not move well.The device broke and the cutter got damaged.It was cut with electrosurgical knife to solve the problem.There was no patient injury.
 
Additional Manufacturer Narrative
H3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the single use loading unit (sulu) was received fully applied, the knife blade was advanced, and the knife blade assembly was damaged.It was reported that the device was fired, but the knife did not transect the tissue and the device broke.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The issue can occur if excessive force was applied to advance the firing knob during application over a thick tissue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the staplers and loading units should not be used on tissue that will compress to less 1.5 mm or 2.0 mm.The staplers and loading units should not be used on tissue that will not comfortably compress to 1.5 mm or 2.0 mm.Such tissue is too thick and closure failure, dehiscence, inability to control bleeding, and tissue trauma could result.The staplers and loading units should not be used to staple liver, spleen and other tissue with sinus beds whose compressibility is such that closure of the instrument may be destructive.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COV LP/LLC, AUGUSTA
1647 perkins rd
augusta GA 30913
MDR Report Key11656876
Report Number1018120-2021-00002
Device Sequence Number3449224
Product Code GDW
UDI-Device Identifier10884521063167
UDI-Public10884521063167
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K885222
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3971
Device Catalogue Number3971
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2021
Initial Date Received by Manufacturer 03/22/2021
Supplement Date Received by Manufacturer05/13/2021
Initial Report FDA Received Date04/13/2021
Supplement Report FDA Received Date06/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-