Model Number 82096-01
Medical Device Problem Codes Break (1069); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: estimated date.It is unknown if the device is returning for analysis.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that an xact carotid stent fractured and failed to deploy.There were no adverse patient effects reported at this time.No additional information was provided regarding this issue.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported break was confirmed by way of the shaft.The reported deployment issue was unable to be tested.There was additional noted bend at the distal sheath likely due to case circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties.It may be possible that the distal shaft was restricted in the anatomy prevented the shaft lumens from moving freely to deploy the stent and as such if the anatomy was challenging may have caused the shaft to separate from the proximal end of the guide wire exit port at the bond; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XACT CAROTID STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11657512
Report Number2024168-2021-03074
Device Sequence Number14237549
Product Code NIM
UDI-Device Identifier08717648010323
UDI-Public08717648010323
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P040038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number82096-01
Device Catalogue Number82096-01
Device Lot Number1011861
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2021
Initial Date Received by Manufacturer 03/24/2021
Supplement Date Received by Manufacturer05/24/2021
Initial Report FDA Received Date04/13/2021
Supplement Report FDA Received Date06/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
-
-