• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L. P. EMPOWR PARTIAL KNEE; EMPOWR PARTIAL KNEETM, TIBIAL BASEPLATE, NP, SIZE 1, LM-RL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L. P. EMPOWR PARTIAL KNEE; EMPOWR PARTIAL KNEETM, TIBIAL BASEPLATE, NP, SIZE 1, LM-RL Back to Search Results
Model Number 357-01-101
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 03/16/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Reported incident - due to patient underwent uka procedure.Post-operative films looked fine.Patient fractured medial tibial plateau soon after surgery, directly below implant keel.Required plating and screws to buttress tibia, correct alignment and induce fracture healing.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was was a revision surgery due to the patient fractured medial tibial plateau.The previous surgery and the surgery detailed in this event occurred 14 days apart.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery.The revision surgery was completed as intended.The device was not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to the patient fractured medial tibial plateau.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions incorrect implant selection, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There were no findings during this evaluation that indicate the reported device was defective.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR PARTIAL KNEE
Common Device Name
EMPOWR PARTIAL KNEETM, TIBIAL BASEPLATE, NP, SIZE 1, LM-RL
Manufacturer (Section D)
ENCORE MEDICAL L. P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key11659510
Report Number1644408-2021-00274
Device Sequence Number11487817
Product Code HSX
UDI-Device Identifier00190446295105
UDI-Public(01)00190446295105
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K191325
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/07/2025
Device Model Number357-01-101
Device Catalogue Number357-01-101
Device Lot Number187Z1018
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/16/2021
Supplement Date Received by Manufacturer04/28/2021
Initial Report FDA Received Date04/13/2021
Supplement Report FDA Received Date05/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age73 YR
-
-