| Model Number |
010118 |
| Medical Device Problem Codes |
Material Twisted/Bent (2981); Component Misassembled (4004)
|
| Health Effect - Clinical Codes |
Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bard 18fr urological catheter was twisted and the guiding nib did not line up with the coude tip.It was stated that it was difficult to ensure the catheter tip was pointing up and there were 17 catheters from a 10 box order.The patient had experienced pain and bleeding.No medical intervention was reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the bard 18fr urological catheter was twisted and the guiding nib did not line up with the coude tip.It was stated that it was difficult to ensure the catheter tip was pointing up and there were 17 catheters from a 10-box order.The patient had experienced pain and bleeding.No medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed as manufacturing related.One sample was confirmed to exhibit the reported failure.Received 9 orange urological catheters unopened in original packaging.3 catheters were selected at random for testing via sampling plan.The first catheter appeared to have a misaligned coude tip.Once the catheter was straightened the coude tip and indicator where aligned.The second and third catheters both had coude indicators that aligned with the tip before and after straightening.The reported event is confirmed, as bends in the first catheter, made the coude indicator appear unaligned with the tip.A potential root cause for this failure could be "operator or machine error.Although there is currently no specification regarding coude tip indicator alignment, the device does not meet specification as the device is intended to have a coude curve aligning with the coude indicator.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review is not required as the reported event was confirmed as manufacturing related.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
|
| |
|