|
Reporter is a j&j employee.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: part number: 388.134.Synthes lot number: 7416645.Supplier lot number: n/a.Release to warehouse date: 11sep2013.Expiration date: n/a.Supplier: tecomet kenosha.No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Service and repair evaluation: it was reported on (b)(6) 2021, during evaluation at service and repair, the torque indicator wrench was found to have failed calibration.The repair technician reported the device failed torque testing.The cause of the issue is failed low.The item will be repaired and will be returned to the customer upon completion of the service and repair process.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|