MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TORQUE INDICATOR WRENCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TORQUE INDICATOR WRENCH Back to Search Results
Model Number 388.134
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is a j&j employee.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported on (b)(6) 2021, during evaluation at service and repair, the torque indicator wrench was found to have failed calibration.There was no patient involvement.This report is for one (1) torque indicator wrench.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: part number: 388.134.Synthes lot number: 7416645.Supplier lot number: n/a.Release to warehouse date: 11sep2013.Expiration date: n/a.Supplier: tecomet kenosha.No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Service and repair evaluation: it was reported on (b)(6) 2021, during evaluation at service and repair, the torque indicator wrench was found to have failed calibration.The repair technician reported the device failed torque testing.The cause of the issue is failed low.The item will be repaired and will be returned to the customer upon completion of the service and repair process.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORQUE INDICATOR WRENCH
Common Device Name
WRENCH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11664720
Report Number2939274-2021-01834
Device Sequence Number3128347
Product Code HXC
UDI-Device Identifier10705034772004
UDI-Public(01)10705034772004
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number388.134
Device Catalogue Number388.134
Device Lot Number7416645
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2021
Initial Date Received by Manufacturer 03/17/2021
Supplement Date Received by Manufacturer04/16/2021
Initial Report FDA Received Date04/14/2021
Supplement Report FDA Received Date05/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-