• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XACT CAROTID STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XACT CAROTID STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number 82087-01
Medical Device Problem Codes Break (1069); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Event or Problem Description
It was reported that the procedure was performed to treat a target lesion in the mildly calcified and moderately tortuous right internal carotid artery.The xact stent delivery system was advanced and some resistance was noted, due to the patient anatomy.The device was removed without difficulty and additional dilatation was performed.During an attempt to re-insert the xact stent delivery system, it was noted that there was a gap at the tip of the device and the nose cone.The device was not used again and there was no adverse patient effect or clinically significant delay.A second xact stent delivery system was used without issue.No additional information was provided.
 
Additional Manufacturer Narrative
Visual analysis was performed on the returned device.The reported difficulty advancing could not be confirmed as it was related to procedural circumstances.The tip damage was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints.The investigation determined that the reported difficulty advancing and damage to the tip was likely related to anatomical conditions.Additionally, the damage to the tip and guide wire lumen likely occurred during the initially failed attempt to advance preventing further use when attempting to reinsert.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XACT CAROTID STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11664911
Report Number2024168-2021-03117
Device Sequence Number3192281
Product Code NIM
UDI-Device Identifier08717648010231
UDI-Public08717648010231
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model Number82087-01
Device Catalogue Number82087-01
Device Lot Number0111361
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2021
Initial Date Received by Manufacturer 03/23/2021
Supplement Date Received by Manufacturer05/06/2021
Initial Report FDA Received Date04/14/2021
Supplement Report FDA Received Date05/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-