| Model Number |
IPN028485 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history review could not be conducted since the lot number was not provided.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
Air leak in the red-blue connector with clamp above connection there is an air leak a with clamp below connection there is no air leak, so it is in the connection.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).A dhr review could not be conducted since the lot number was not provided.A visual inspection of the product involved in the complaint was performed on only the ats connectors received from the customer under customer complaints (b)(4) of product code s-1102-08lf (pe sahara dry suct/dry seal dual lf 6).However, ats connectors provided by the customer were received with the tubing cut at the base of the connectors as opposed to the requested 6 inches of tubing on both sides of the connector.No additional damage was found on the received sample that can lead to a leaking issue.Also, it was confirmed that the tubing is assembled correctly on each end of the ats connectors.The assembly of the o-ring was verified, and it was found assembled correctly.Functional inspection test cannot be performed since the sample involved in this customer complaint was receive incomplete only the ats connectors were received for evaluation.Received sample was connected to a bottle to simulate the use using water; it was tested with suction and no leakage issues were observed, the sample was connected and disconnected several times to see if any of these disconnections could lead to a leakage issue and no leaks were observed.A visual inspection of the product involved in the complaint was performed on only ats connectors received from the customer under customer complaints (b)(4) of product code s-1102-08lf (pe sahara dry suct/dry seal dual lf 6).However, ats connectors provided by the customer were received with the tubing cut at the base of the connectors as opposed to the requested 6 inches of tubing on both sides of the connector.No additional damage was found on the received sample that can lead to a leaking issue.Also, it was confirmed that the tubing is assembled correctly on each end of the ats connectors.The assembly of the o-ring was verified, and it was found assembled correctly.(see attached pictures) received sample was connected to a bottle to simulate the use using water; it was tested with suction and no leakage issues were observed, the sample was connected and disconnected several times to see if any of these disconnections could lead to a leakage issue and no leaks were observed; based on sample received and testing perform to it, customer complaint cannot be confirmed; nuevo laredo facility will continue to track and trend this failure mode.Based on previous statements it is not possible to establish corrective actions.
|
| |
|
Event or Problem Description
|
|
Air leak in the red-blue connector with clamp above connection there is an air leak a with clamp below connection there is no air leak, so it is in the connection.
|
| |
|
Search Alerts/Recalls
|