• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION UNKNOWN VENA CAVA FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION UNKNOWN VENA CAVA FILTER FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKFILTER
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Coagulation Disorder (1779); Perforation (2001); Perforation of Vessels (2135); Stenosis (2263); Thrombosis/Thrombus (4440)
Event Date 08/26/2020
Event Type  Injury  
Manufacturer Narrative
As reported, a patient underwent placement of an unknown cordis inferior vena cava (ivc) filter. The indication for filter placement is not available. The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to filter perforation, ivc stenosis and caval thrombosis. The filter remains implanted; thus, unavailable for analysis. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The cordis vena cava filters are indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event. A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters. Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place. Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation. Stenosis of the ivc is associated with all ivc filter products and does not represent a device malfunction. A protective inferior vena cava (ivc) filter may later be incorporated into a chronic post-thrombotic ilio-caval obstruction (occlusive, requiring recanalization, or nonocclusive). Obstruction of varying types of ivc filters may occur due to primary thrombosis of the filter or capture of large emboli. Permanent ivc filters have been reported to obstruct in up to 20% of patients. Blood clots and occlusive thrombosis within the filter and vasculature do not represent a device malfunction. Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics. Without procedural films or images for review the reported event(s) could not be confirmed. Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal brief, a patient underwent placement of an unspecified vena cava filter manufactured by cordis. The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to filter perforation, inferior vena cava (ivc) stenosis and caval thrombosis. As a direct and proximate result, the patient suffered life-threatening injuries and damages and required extensive medical care and treatment. As a further proximate result, the patient has suffered and will suffer significant medical expenses, pain and suffering and other damages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN VENA CAVA FILTER
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014 2802
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014 2802
Manufacturer Contact
karla castro
14201 nw 60th avenue
miami lakes, FL 33014-2802
7863138372
MDR Report Key11665796
MDR Text Key245937732
Report Number9616099-2021-04409
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue NumberUNKFILTER
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/14/2021 Patient Sequence Number: 1
-
-