• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER Back to Search Results
Model Number 995081
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the patients were assigned to perform percutaneous nephrolithotomy.Before use, they were all checked that the outer package of the products was complete and there was no abnormality found.After opening the outer package, it was found that the internal disinfection package was not sealed.
 
Additional Manufacturer Narrative
The reported event was confirmed as supplier related.The root cause for this event would be "operator error missed sealing pouch".Photo samples were provided by the customer in addition to the returned sample.The photo samples show the unsealed pouch.The physical sample was evaluated.Some compression damage was noted on the corners of the carton, possibly due to storage of the product.The sample was removed from the carton and the tyvek pouch was not sealed on one end; however, the chevon and length seals were still intact, and all contents were still in the tray.The returned sample packaging did not appear to have markings to indicate that it was sealed with robot sealer.No pattern or impression marks from heating dye on either the clear film or the tyvek bottom.There were no tears in the packaging or any other indications that external force was applied to result in it being open in the carton.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review was not required as a review of the label could not have prevented the reported event.Correction: h.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the patients were assigned to perform percutaneous nephrolithotomy.Before use, they were all checked that the outer package of the products was complete and there was no abnormality found.After opening the outer package, it was found that the internal disinfection package was not sealed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER
Common Device Name
BALLOON DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11665849
Report Number1018233-2021-02125
Device Sequence Number2935726
Product Code LJE
UDI-Device Identifier00801741080715
UDI-Public(01)00801741080715
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063632
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2021
Device Model Number995081
Device Catalogue Number995081
Device Lot NumberBMDPFM13
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/25/2021
Supplement Date Received by Manufacturer12/16/2021
Initial Report FDA Received Date04/14/2021
Supplement Report FDA Received Date12/17/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-