MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 50MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 50MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Model Number 186731750
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: additional device product codes: (b)(4).Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, the surgeon used custom awls to access the pedicle, and then inserted the 7.0x50mm cortical fix screw at l5 on the right side.As the screw was being inserted the tulip head popped off disengaging from the body of the screw.The surgeon attempted to remove the body of the screw with a t20 driver but stripped the t20 cannulated driver.The surgeon also stripped the t20 non cannulated driver and fractured the tip of the mini open driver.The broken tip of the mini open driver was possibly inside the hex of the screw.The next attempt to remove the shaft of the screw was to use a screw removal reverse thread instrument, which failed to remove the screw.The surgeon was able to successfully remove the fractured instrument and confirmed with both x-ray and matching it up, that the instrument was removed.The screw remained in the patient with a rescue screw 5.0x50mm placed directly beside it.To insert the rescue screw, the surgeon used a jam shidi, and guide wire with 4.35mm tap.The rescue screw had no issue, and the surgeon was able to pass the rod and finish up.The tulip was recovered, the shaft of the 7.0x50mm screw remains in the patient, the tip of the mini hex driver is missing, and the fragment from the screw removal set was removed and accounted for.There was a surgical delay of estimated thirty (30) to forty-five (45) minutes.Patient status is unknown.The procedure was successfully completed.This report involves one (1) viper system cortical fix polyaxial screw 5.5 7 x 50mm.This is report 1 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: a product investigation was conducted.Visual inspection: the 5.5 ti cort fix 7x50mm (part#: 186731750, lot#: 198093) was returned and received at us cq.Upon visual inspection, it was observed that the rod was broken off the head and the broken piece was not returned to the us cq.The embedded piece cannot be confirmed as there was no evidence (photos or x-rays) provided showing that the fragments were embedded in the patient.No other issues were identified with the returned device.Document/specification review: based on the date of manufacture the relevant drawings, the current and manufactured revision of drawings were reviewed.Dimensional inspection: the dimensional inspection was not performed due to no access to relevant components.Investigation conclusion: the complaint condition is confirmed for the 5.5 ti cort fix 7x50mm (part#: 186731750, lot#: 198093) as the device received was broken.However, the embedded device allegation was not confirmed as there were no photos or x-rays provided.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.The potential cause could be due to unintended forces applied to the device.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: a device history record (dhr) review was conducted: part#: 186731750, lot#: 198093.The dhr of product code: 186731750, lot: 198093, was reviewed and no non-conformances were observed during the manufacturing process.The product was released on may 18, 2018.Qty.99.The dhr was electronically reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 7 X 50MM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key11666047
Report Number1526439-2021-00715
Device Sequence Number16926136
Product Code MNH
UDI-Device Identifier10705034352022
UDI-Public(01)10705034352022
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K110216
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186731750
Device Catalogue Number186731750
Device Lot Number198093
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Initial Date Received by Manufacturer 03/17/2021
Supplement Date Received by Manufacturer04/19/2021
Initial Report FDA Received Date04/14/2021
Supplement Report FDA Received Date05/13/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REMOVAL INSTRUMENT.; T20 SCREWDRIVER SHAFT.; UNKNOWN AWLS.; UNKNOWN GUIDE/COMPRESSION/K-WIRES.; UNKNOWN RODS.; UNKNOWN SPD PRODUCT.; VIPER UNIVERSAL POLY DRIVER.; VIPER2 T20 HEXLOBE DRIVER SHAF.
Patient Age59 YR
-
-