• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN LNR CON +4 10D 28IDX50OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN LNR CON +4 10D 28IDX50OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS Back to Search Results
Model Number 1218-28-750
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Fall (1848); Joint Dislocation (2374)
Date of Event 03/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient had original surgery (b)(6) 2021.Patient fell and had a dislocation.Revised liner to constrained liner on (b)(6) 2021.Patient then fell out of bed causing second dislocation.Second revision completed today.Doctor removed and re-implanted stem, constrained liner and head ball.
 
Event or Problem Description
Additional information was received stating that the reason the previous stem was revised was because it had subsided.There was no surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Provided x-ray images have been reviewed.Although a full dislocation event is not depicted there is an appearance of the femoral head having moved downward and appears partially dislocated from the constrained liner.It is not possible to determine the root cause using only x-ray images.The patient sustained a fall and this may have been a factor.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN LNR CON +4 10D 28IDX50OD
Common Device Name
PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11667235
Report Number1818910-2021-07946
Device Sequence Number14224076
Product Code KWZ
UDI-Device Identifier10603295011538
UDI-Public10603295011538
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K043058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1218-28-750
Device Catalogue Number121828750
Device Lot NumberJ00M71
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/30/2021
Supplement Date Received by Manufacturer04/19/2021
05/10/2021
Initial Report FDA Received Date04/14/2021
Supplement Report FDA Received Date04/29/2021
05/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DLT TS CER HD 12/14 28MM +1.5; PINN LNR CON +4 10D 28IDX50OD; SUMMIT POR TAPER SZ4 STD OFF
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
-
-