MAUDE Adverse Event Report: VASCULAR SOLUTIONS, LLC RAIDER GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VASCULAR SOLUTIONS, LLC RAIDER GUIDEWIRE Back to Search Results
Model Number 7448
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hematoma (1884); Perforation (2001)
Date of Event 01/13/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing records will be reviewed.A follow-up report will be issued after the investigation is complete.
 
Event or Problem Description
As reported: during a successful pci to a non-tortuous, moderately calcific cto of the proximal rca on cto pci study subject (b)(6), subject suffered an ellis grade 1 vessel perforation to an epicardial collateral from the cfx which resulted in a residual hematoma.Many devices, both study and non-study, were used in the procedure.Study devices included a turnpike microcatheter, turnpike lp microcatheter and a raider guidewire.It is unknown which device caused the perforation but felt to be caused by a non-study device.Perforation treated with coil embolization (0.014" coils x 2) with resolution, hematoma treated with cutting balloon with moderate success of decompression and small residual hematoma.Tee performed prior to discharge showed no pericardial effusion.Discharged home next day with no further sequelae.Associated to mdrs: 2134812-2021-00033 turnpike smooth.2134812-2021-00034 turnpike lp.
 
Additional Manufacturer Narrative
There was no product returned for this complaint.No returned product evaluation could be completed.Lot 680191 was reviewed.There are no nonconformances related to this lot therefore supporting the device met material, assembly and performance specifications.Case details was reviewed.A patient underwent a pci procedure.Vessel was moderately calcific and non-tortuous.The patient endured an ellis grade 1 perforation to an epicardial collateral from the cfx which resulted in residual hematoma.A raider (study device) was one of the devices used in the procedure.Per ifu, a vessel perforation and hematoma are identified as potential adverse effects that may be associated with the use of the raider guidewires.No product was returned to vsi/teleflex for investigation.It is unknown if any damages were present on the raider that could have contributed to the event.Multiple ancillary devices were also used in the case.The event details states that it is unknown which device caused the perforation but likely a non-study device.The perforation was treated with coil embolization (0.014" coils x 2).Hematoma treated with cutting balloon with moderate success of decompression.Tee performed prior to discharge showed no pericardial effusion.Patient discharged home next day with no further sequelae.Based on the information, the most likely root cause of the issue is undeterminable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAIDER GUIDEWIRE
Common Device Name
GUIDEWIRE
Manufacturer (Section D)
VASCULAR SOLUTIONS, LLC
6464 sycamore court north
minneapolis MN 55369
MDR Report Key11668191
Report Number2134812-2021-00032
Device Sequence Number2936400
Product Code DQX
UDI-Device Identifier10841156101275
UDI-Public(01)10841156101275
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K173532
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/04/2022
Device Model Number7448
Device Catalogue Number7448
Device Lot Number680191
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/25/2021
Supplement Date Received by Manufacturer03/25/2021
Initial Report FDA Received Date04/14/2021
Supplement Report FDA Received Date07/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age70 YR
-
-