Model Number PXSLIMLAN115STR
Medical Device Problem Codes Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2021-00757.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the abdominal aortic aneurysm (aaa) sac using pod packing coils (pod pcs), a lantern delivery microcatheter (lantern), and non-penumbra diagnostic catheter.During the procedure, the physician placed two pod pcs in the target location.Subsequently, the next pod pc became stuck inside the lantern after advancing to a certain point and would not retract.It was reported the physician had advanced the diagnostic catheter without a wire and had manipulated the lantern without a wire.When attempting to pull the pod pc back, the pod pc unintentionally detached in the lantern.Therefore, the detached pod pc and lantern were removed together.The lantern was inspected, and the hospital staff noticed the lantern was kinked at the distal shaft, around the area where the pod pc was getting stuck.The procedure was completed using two new pod pcs and a new lantern.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11668439
Report Number3005168196-2021-00756
Device Sequence Number2980570
Product Code DQY
UDI-Device Identifier00814548016610
UDI-Public00814548016610
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K152840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/19/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/05/2023
Device Model NumberPXSLIMLAN115STR
Device Catalogue NumberPXSLIMLAN115STR
Device Lot NumberF100831
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2021
Initial Report FDA Received Date04/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80
Date Report Sent to FDA01/01/2005
-
-