Model Number 7209016
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during a surgery, the biorci screw broke while it was being deployed.The broken pieces were removed by using tweezers.The procedure was successfully completed without a significant delay using a back-up device.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the polymer found that the storage protocols, material specifications, and material tests were appropriately documented.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H3, h6: the device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.The device was returned in the original packaging.The distal tip of the screw is broken.Based on the condition of the product material found during visual inspection it was determined the device damage was caused by the application of improper/excessive force.Additional material testing is not required.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the polymer found that the storage protocols, material specifications, and material tests were appropriately documented.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCR BIORCI-HA 10X25 STERILE
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11668451
Report Number1219602-2021-00861
Device Sequence Number3085141
Product Code HWC
UDI-Device Identifier03596010460035
UDI-Public03596010460035
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K002274
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/02/2025
Device Model Number7209016
Device Catalogue Number7209016
Device Lot Number50866528
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2021
Initial Date Received by Manufacturer 03/24/2021
Supplement Date Received by Manufacturer05/12/2021
07/19/2021
Initial Report FDA Received Date04/14/2021
Supplement Report FDA Received Date05/13/2021
07/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-