STRYKER GMBH LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L18MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 657318
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device is not available; disposed by hospital.
 
Event or Problem Description
As reported: "variax 2 t10 locking screws were not engaging with the distal cluster holes in the anatomic fibula plates.4 screws in a row.It happened while trying to lock the screws to the plate.Prolongation of the case due to inserting, not engaging, removing, drilling again, measuring again and implanting a shorter locking screw to only have it not engage again.".
 
Additional Manufacturer Narrative
The reported event that the locking screw variax2 t10, full thread, 3.5mm / l16mm was alleged of 'implant - insertion impaired' could not be confirmed since the device was not returned for evaluation and no other evidence was provided.A device inspection was not possible since the affected device was not returned and no other evidences were provided for investigation.A potential non-conformity report was initiated to address similar events in the past.The comprehensive root cause analysis of the potential ncr included a surface and microstructure analysis as well as nano-indent hardness measurement of variax1 and variax2 screw samples by the fraunhofer institute on behalf of stryker.Further internal investigation efforts were focusing on the anodization color difference between variax1 and variax2 screws.Significant differences between the screws which could result in the reported complaints have not been revealed in either fraunhofer or stryker investigations.Excessive in-house handling and bench testing with variax2 screws from various manufacturing batches showed that locking is achieved as intended.The desired increase in torque indicating to the customer that the screw is locking in the plate, can be confirmed.The batch record could not be reviewed because the affected device was not returned, and the lot number was not communicated.A review of the labeling did not indicate any abnormalities.Considering the information given and based on the above investigations a root cause of the reported event could not be determined.If any further substantial information is provided, the investigation report will be updated.
 
Event or Problem Description
As reported: "variax 2 t10 locking screws were not engaging with the distal cluster holes in the anatomic fibula plates.4 screws in a row.It happened while trying to lock the screws to the plate.Prolongation of the case due to inserting, not engaging, removing, drilling again, measuring again and implanting a shorter locking screw to only have it not engage again.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L18MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key11668562
Report Number0008031020-2021-00151
Device Sequence Number3049441
Product Code HWC
UDI-Device Identifier07613327086911
UDI-Public07613327086911
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180500
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number657318
Device Catalogue Number657318
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/17/2021
Supplement Date Received by Manufacturer04/26/2021
Initial Report FDA Received Date04/14/2021
Supplement Report FDA Received Date05/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-