• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA CAPIOX CUSTOM PACK; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA CAPIOX CUSTOM PACK; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number CX-XRY54602
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2021
Event Type  malfunction  
Manufacturer Narrative
Udi - not required for product code.Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation-clinical engineer.Pma/510(k)- k130520.The actual sample was received for evaluation.Visual inspection revealed that the lock adapter had come off the male connector of the sampling system.Magnifying inspection of the actual sample did not find any visible anomaly, including deformity, in the male connector or in the lock adapter.The outer diameter of the rib of the male connector and the inner diameter of the lock adapter were measured.Compared with a current product sample, no difference was observed in the measured values.The surface of the male connector was subjected to elemental analysis by sem-edx (scanning electron microscope / energy dispersive x-ray spectroscopy).The result showed the presence of si, which is likely to have been derived from the silicone applied to the switch cocks of the three-way stopcock to improve the lubricity of them in the manufacturing process.Reproductive testing was performed, and silicone was applied to a male connector of a factory-retained sampling system, a female connector was connected to it, and then a lock adapter was tightened up.As a result, the lock adapter came off.A review of the device history record and product-release judgement record of the involved product code/lot# combination was conducted with no findings.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.It is likely that the silicon, which is applied to the switch cocks of the sampling system to improve the lubricity, was transferred to the male connector part due to some factors.Afterward, the lock adapter, when re-tightened, may have got over the rib of the male connector in lubricated state and came off.However, the exact cause of the reported event cannot be definitively determined based on the available information.(b)(4).
 
Event Description
The user facility reported that the capiox custom pack was used pre-treatment.Before priming, when they attempted to tighten the lock adapter of the sampling system, it came off.They replaced the sampling system with a triple stopcock from top to cope with the problem.The patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX CUSTOM PACK
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
mary o'neill
reg. no. 2243441
265 davidson ave suite 320
somerset, NJ 08873
8002837866
MDR Report Key11670145
MDR Text Key245537509
Report Number9681834-2021-00060
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071494
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Catalogue NumberCX-XRY54602
Device Lot Number200623
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/31/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-