Model Number PXSLIMLAN150STR
Medical Device Problem Codes Failure to Advance (2524); Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the superior mesenteric artery (sma) to the inferior mesenteric artery (ima) using a lantern delivery microcatheter (lantern), pod packing coils (pod pcs), a non-penumbra catheter.During the procedure, while inserting the lantern into the catheter the physician encountered resistance and subsequently, the lantern became kinked.The physician attempted to advance a pod pc through the lantern and was unable to; therefore, the pod pc was re-sheathed and the lantern was removed from the patient.The procedure was completed using a new lantern, the same pod pc, additional pod pcs, and the same catheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11671366
Report Number3005168196-2021-00759
Device Sequence Number2982376
Product Code DQY
UDI-Device Identifier00814548016672
UDI-Public00814548016672
Combination Product (Y/N)Y
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K152840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/19/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/09/2023
Device Model NumberPXSLIMLAN150STR
Device Catalogue NumberPXSLIMLAN150STR
Device Lot NumberF100101
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2021
Initial Report FDA Received Date04/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
Date Report Sent to FDA01/01/2005
-
-