• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS SMALL 3.5MM DEPTH GAUGE LONG; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS SMALL 3.5MM DEPTH GAUGE LONG; PLATE, FIXATION, BONE Back to Search Results
Model Number 71175068
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2021
Event Type  malfunction  
Event Description
It was reported that, during surgery, it was noticed that the tip of the inner sleeve on the depth gauge had fallen off.The procedure was successfully completed without delay using a smith+nephew back-up device.No patient injury or other complications were reported.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The tip of the device was damaged rendering it inoperable.The device show signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Additional information: d4, g1, h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL 3.5MM DEPTH GAUGE LONG
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11671377
MDR Text Key245580458
Report Number1020279-2021-02987
Device Sequence Number1
Product Code HRS
UDI-Device Identifier00885556635865
UDI-Public00885556635865
Combination Product (y/n)N
PMA/PMN Number
K162078
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71175068
Device Catalogue Number71175068
Device Lot Number17MTRS0007
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-