Model Number FX649T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 03/13/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation on- going.Additional information/ investigation results will be provided in a supplemental report.
 
Event or Problem Description
It was reported that a progav 2.0 shuntsystem ((b)(4) ) was implanted during a procedure performed on (b)(6) 2020.According to the complainant, the shuntsystem was believed to be operated in over- drainage.The ventricle are narrowed.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Patient information: (b)(6).Gender: female.
 
Additional Manufacturer Narrative
Investigation: visual inspection: no significant deformations or damage of the valves were detected during the visual inspection.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the progav 2.0 is operating within the accepted tolerance in the horizontal position.The shuntassistant 2.0 is operating not within the specified tolerances in the vertical position.The shuntassistant 2.0 showed a slightly accelerated outflow.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Results: first, we performed a visual inspection of the progav 2.0 shunt system.No significant deformations or damage of the valves were detected during the visual inspection.Next, we tested the permeability, adjustability, and opening pressure of the valve, as well as the brake functionality and brake force.The opening pressure of the shuntassistant 2.0 was out of tolerance, but all other specifications were met.Finally, we have dismantled the valves.Inside both valves we have found build-up of substances (likely protein).Based on our investigation, we confirm that the shuntsystem shows an accelerated flow of liquid at the time of our investigation.This is likely due to the deposits observed inside the valves.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 W/SA2.0 25 & DISTAL CATHETER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
MDR Report Key11671467
Report Number3004721439-2021-00080
Device Sequence Number2988060
Product Code JXG
UDI-Device Identifier04041906502072
UDI-Public4041906502072
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K190174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date02/05/2025
Device Model NumberFX649T
Device Catalogue NumberFX649T
Device Lot Number20045905
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2021
Supplement Date Received by Manufacturer04/26/2021
Initial Report FDA Received Date04/15/2021
Supplement Report FDA Received Date05/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age14 YR
Patient Weight50
-
-