| Model Number |
NX029Z |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported to (b)(6) inc.That a as vega ps femoral comp.Cemented f4n r (part # nx029r) was implanted during a primary total knee arthroplasty (tka) procedure performed on (b)(6), 2018.According to the complainant, the device loosened 2 years and 6 months post surgery.The patient underwent a revision surgery.Upon removal, the femur exhibited very little cement bonding.The complaint device was discarded, therefore, it is not available to be returned to the manufacturer for evaluation.A revision surgery was necessary.The adverse event is filed under aag reference (b)(4).Involved components: nx110 - vega ps gliding surface t1/1+ 10mm - lot 52396660.Nx051z - as vega ps tibial plateau cemented t1 - lot 52424675.Nn261z - as tibial obturator d12mm - lot 52413272.
|
| |
|
Event or Problem Description
|
|
The adverse event is filed under aag reference (b)(4) ((b)(4)).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Explanation and rationale, and root cause: according to the quality standard and device history records (dhr) a material defect and production error was not found.In the light of the small amount of information received and due to the circumstance that we did not receive the complained devices for investigation, it is not possible to determine a definitive root cause for the mentioned failure.It could be possible that there were problems with the cement technique.Corrective action for the topic of vega loosening, a product safety case (psc) was initiated.
|
| |
|