| Model Number |
MN60AC |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that during a cataract surgery with an intraocular lens (iol) implantation, the doctor noticed that the optic was stained with unknown material upon inserting the iol into the eye.The doctor decided to remove and replace the iol with a backup iol.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was returned in the lens case.Viscoelastic was observed on the lens.The lens was intact with no optic or haptic damage observed.It is unclear if the lens was implanted a removed, as no damage was observed.The root cause for the reported "stain" could not be determined.The lens was returned undamaged with only dried viscoelastic observed.Information was provided that the surgeon noted, the stain on the optic area and flushed out.The surgeon decided not to use the lens.And replaced it with a back up lens.The manufacturer internal reference number is: (b)(4).
|
| |
|