• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number MN60AC
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that during a cataract surgery with an intraocular lens (iol) implantation, the doctor noticed that the optic was stained with unknown material upon inserting the iol into the eye.The doctor decided to remove and replace the iol with a backup iol.
 
Additional Manufacturer Narrative
The lens was returned in the lens case.Viscoelastic was observed on the lens.The lens was intact with no optic or haptic damage observed.It is unclear if the lens was implanted a removed, as no damage was observed.The root cause for the reported "stain" could not be determined.The lens was returned undamaged with only dried viscoelastic observed.Information was provided that the surgeon noted, the stain on the optic area and flushed out.The surgeon decided not to use the lens.And replaced it with a back up lens.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF MULTIPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514979
MDR Report Key11673356
Report Number1119421-2021-00828
Device Sequence Number11552527
Product Code HQL
UDI-Device Identifier00380655095201
UDI-Public00380655095201
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Model NumberMN60AC
Device Catalogue NumberMN60AC.245
Device Lot Number12491435
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer10/29/2021
Initial Report FDA Received Date04/15/2021
Supplement Report FDA Received Date11/24/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DISCOVISC OPHTHALMIC VISCOSURGICAL DEVICE; MONARCH II IOL DELIVERY SYSTEM, CARTRIDGE B; MONARCH III IOL DELIVERY SYSTEM, INJECTOR
Patient SexMale
-
-