MAUDE Adverse Event Report: ETHICON INC. PROLENE SUT 30IN(75CM) 6-0 BLU; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PROLENE SUT 30IN(75CM) 6-0 BLU; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Model Number 8776H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Component code: device not returned.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The single complaint was reported with multiple events.There are no additional details regarding the additional events.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: what was being done when the needle came loose from the suture (removal from package / during passage through tissue/ during tying? was the needle removed from the patient during the same procedure? unknown.What tissue/location was suturing when pull-off occurred? unknown.Please provide the lot number for the involved device.Qmbcqd.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.
 
Event or Problem Description
It was reported that a patient underwent a vascular procedure in 2021 and suture was used.During the procedure, the suture broke off from the needle like a pop off.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).Date sent to fda: 05/07/2021.Additional information: d9, h3, h6.H3 analysis summary: visual analysis of the returned sample determined that two sutures¿ pieces of product code 8776h were received for evaluation the sutures ends were noted with marks that appears to be by use of a surgical instrument.The needle was not received for evaluation.No conclusion could be reached as to what caused the reported complaint.Additionally, visual analysis of the returned sample determined that eighteen unopened samples of product code 8776h were received for evaluation.Visual inspection of unopened samples and no defects were found on the package.The samples were opened, and the swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand and no defects or damage were observed.A functional test was performed using an instron and the pull force was above the minimum requirements.The device performed without any defect noted and the actual complaint sample was not returned for analysis.Related reports: 2210968-2021-03405.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE SUT 30IN(75CM) 6-0 BLU
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
MDR Report Key11673596
Report Number2210968-2021-03407
Device Sequence Number3196505
Product Code GAW
UDI-Device Identifier10705031020481
UDI-Public10705031020481
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model Number8776H
Device Catalogue Number8776H
Device Lot NumberQMBCQD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2021
Initial Date Received by Manufacturer 03/22/2021
Supplement Date Received by Manufacturer04/12/2021
Initial Report FDA Received Date04/15/2021
Supplement Report FDA Received Date05/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-